饮用水硝酸盐氮的测定
饮用水硝酸盐氮的测定
摘要 :用紫外分光光度法测定饮用水、地下水中的硝酸盐氮,在0~4mg/L范围内硝酸盐氮的浓度与在220nm 、275nm的吸光度呈线性关系,相关系数r=0.9998。测定结果的相对标准偏差为0.2%,检出限为0.023mg/L. 与 酚二磺酸光度法进行比对,经F和t检验,两种方法无显著性差异。
关键词:紫外分光光度法;饮用水;硝酸盐氮
Abstract: With ultraviolet spectrophotometry to test the nitrate nitrogen in drinking water and ground water, it concludes that in 0~4mg/L, the concentration of nitrate nitrogen presents linear relation with it’s absorbance in 220nm and 275nm, and related coefficient of r=0.9998. The relative standard deviation of the test results is 0.2%, and detection limit is 0.023mg/L. by the detection of F and t, the ultraviolet spectrophotometry has no significant difference with phenol disulfonic acid spectrophotometry.
Key words: ultraviolet spectrophotometry; drinking water; nitrate nitrogen
水中的硝酸盐是在有氧环境下,亚硝氮、氨氮等各种形态的含氮化合物中最稳定的氮化合物,亦是含氮有机化合物无机化作用的最终分解产物。亚硝酸盐可氧化而生成硝酸盐,硝酸盐在无氧的环境中,亦可受微生物作用而还原为亚硝酸盐。水中硝酸盐氮含量相差悬殊,从10ug/L至10mg/L,情节的地表水含量较低,受污染水体,以及一些深层地下水中含量较高,摄入硝酸盐后,经肠道中微生物作用转变成亚硝酸盐而出现毒性作用。文献报道,水中硝酸盐含量达数十毫克/升时,可致婴儿中毒。通常用酚二磺酸光度法测定饮用水、地下水中的硝酸盐氮,该方法:试剂稳定性差、操作繁琐、容易照成实验室内污染、线性较差。笔者采用的紫外分光光度法测定硝酸盐,克服了传统方法的缺点,操作简单、干扰少、试剂用量少,现行较好等优点。
1、方法原理
利用硝酸根离子在220nm波长处的吸收而测定硝酸盐氮。溶解的有机物在220nm处也会有吸收,而硝酸根离子在275nm处没有吸收。因此,在275nm处另做一次测量,以校正硝酸盐氮的值。
2、实验部分
2.1仪器:TU—1810紫外分光光计
离子交换柱(Φ1.4cm 装树脂高度5-8cm)
2.2试剂:氢氧化铝悬浮液